Measuring principle | Enzymatic - amperometric |
Sensor | BST Thick film biosensor |
Response time | Approx. 5 s |
Measuring range | 0,5 - 33,3 mmol/l (9 - 600 mg/dl) |
Serial imprecision | VK 5 % bei 5 mmol/l (capillary blood, n=10) |
Sensor lifetime | 30 days |
Sample | Capillary whole blood |
Sample volume | Approx. 6 _6;l |
Control samples | Internal management of 2 BST control samples |
Internal check of instrument | RiLiBÄK conform |
Working temperature | +15°C bis +35°C |
Storage temperature | -20°C bis +70°C (without Sensor) |
Humidity | 20 % bis 95 % |
Memory | 600 data (first in - first out) |
Dimensions (L x W x H) | Approx. 163 x 122 x 38 mm |
Weight | Approx. 500 g |
Power supply | 230 V - 50/60 Hz |
Power consumption | max. 9 VA |
Options | Docking station, scanner, wireless external printer, HIS/LIS networking |